Revisiting Approved Digital Biomarkers: A Context-of-Use Perspective on Evidence Design
DOI:
https://doi.org/10.51094/jxiv.6720キーワード:
Digital Biomarker、 V3 Framework、 Validation、 Context of Use抄録
Background
Digital biomarkers (dBMs) with adequate sensor accuracy and algorithm performance do not necessarily reach implementation. We examined whether the barrier lies less in the quantity of validation evidence than in a mismatch with the evidence a dBM's context of use (CoU) demands.
Summary
Using the V3+ framework of the Digital Medicine Society (DiMe) as a diagnostic lens, we retrospectively mapped the public regulatory record of five purposively selected cases: stride velocity 95th centile (SV95C), the Apple Watch atrial fibrillation (AF) History Feature, FreeStyle Libre, Mobilise-D digital mobility outcomes (DMOs), and 24-h cough frequency in chronic cough. No single model of success emerged; the load-bearing requirement differed by CoU and by implementation pathway. The load-bearing requirement was agreement with the reference method and usability for the two device cases, and a narrowly specified CoU for SV95C. The Mobilise-D DMOs remain at a letter of support, with cross-disease transferability and the minimal important difference (MID) still to be established; for cough frequency, the contested requirement shifted from validation of the counting method to the clinical meaningfulness of the change.
Key Messages
V3+ is most useful not as a uniform pass/fail scorecard but as a diagnostic lens that reveals which requirement is load-bearing in a given CoU. Because the CoU granted for the same device can differ across markets, and the evidence demanded follows the CoU, setting the target regulatory authority and the CoU is the first and most consequential decision in evidence design.
利益相反に関する開示
The authors' institutional affiliations are listed on the title page. During the three years preceding submission, no author received any grant, consultancy fee, honorarium, speaker fee, travel support, stock or stock options, patent royalty, or other finダウンロード *前日までの集計結果を表示します
引用文献
Di J, Demanuele C, Kettermann A, Karahanoglu FI, Cappelleri JC, Potter A, et al. Considerations to address missing data when deriving clinical trial endpoints from digital health technologies. Contemp Clin Trials. 2022;113:106661.
Babrak LM, Menetski J, Rebhan M, Nisato G, Zinggeler M, Brasier N, et al. Traditional and Digital Biomarkers: Two Worlds Apart? Digit Biomark. 2019;3(2):92–102.
Erdemli G, Murphy T, Walinsky S. Regulatory considerations for successful implementation of digital endpoints in clinical trials for drug development. Npj Digit Med. 2025;8(1):142.
Tackney MS, Steele A, Newman J, Fritzsche M-C, Lucivero F, Khadjesari Z, et al. Digital endpoints in clinical trials: emerging themes from a multi-stakeholder Knowledge Exchange event. Trials. 2024;25(1):521.
Bakker JP, Izmailova ES, Clement A, Hoffmann S, Leptak C, Menetski JP, et al. Regulatory Pathways for Qualification and Acceptance of Digital Health Technology-Derived Clinical Trial Endpoints: Considerations for Sponsors. Clin Pharmacol Ther. 2025;117(1):56–72.
Aboagye NY, Hinchliffe C, Del Din S, Ng W-F, Baker KF, Baker MR. Systematic review: digital biomarkers of fatigue in chronic diseases. Npj Digit Med. 2025;8(1):602.
Macias Alonso AK, Hirt J, Woelfle T, Janiaud P, Hemkens LG. Definitions of digital biomarkers: a systematic mapping of the biomedical literature. BMJ Health Care Inform. 2024;31(1):e100914.
Goldsack JC, Dowling AV, Samuelson D, Patrick-Lake B, Clay I. Evaluation, Acceptance, and Qualification of Digital Measures: From Proof of Concept to Endpoint. Digit Biomark. 2021;5(1):53–64.
Izmailova ES, Wagner JA, Perakslis ED. Wearable Devices in Clinical Trials: Hype and Hypothesis. Clin Pharmacol Ther. 2018;104(1):42–52.
Bakker JP, Barge R, Centra J, Cobb B, Cota C, Guo CC, et al. V3+ extends the V3 framework to ensure user-centricity and scalability of sensor-based digital health technologies. Npj Digit Med. 2025;8(1):51.
Goldsack JC, Coravos A, Bakker JP, Bent B, Dowling AV, Fitzer-Attas C, et al. Verification, analytical validation, and clinical validation (V3): the foundation of determining fit-for-purpose for Biometric Monitoring Technologies (BioMeTs). Npj Digit Med. 2020;3(1):55.
Manta C, Patrick-Lake B, Goldsack JC. Digital Measures That Matter to Patients: A Framework to Guide the Selection and Development of Digital Measures of Health. Digit Biomark. 2020;4(3):69–77.
Servais L, Strijbos P, Poleur M, Mirea A, Butoianu N, Sansone VA, et al. Evidentiary basis of the first regulatory qualification of a digital primary efficacy endpoint. Sci Rep. 2024;14(1):29681.
European Medicines Agency. Final qualification opinion for stride velocity 95th centile as primary endpoint in studies in ambulatory Duchenne muscular dystrophy studies (EMADOC-1700519818-1127132) [Internet]. 2023 [cited 2026 May 21]. Available from: https://www.ema.europa.eu/en/documents/scientific-guideline/qualification-opinion-stride-velocity-95th-centile-primary-endpoint-studies-ambulatory-duchenne-muscular-dystrophy-studies_en.pdf
Servais L, Eggenspieler D, Poleur M, Grelet M, Muntoni F, Strijbos P, et al. First regulatory qualification of a digital primary endpoint to measure treatment efficacy in DMD. Nat Med. 2023;29(10):2391–2.
Servais L, Yen K, Guridi M, Lukawy J, Vissière D, Strijbos P. Stride Velocity 95th Centile: Insights into Gaining Regulatory Qualification of the First Wearable-Derived Digital Endpoint for use in Duchenne Muscular Dystrophy Trials. J Neuromuscul Dis. 2022;9(2):335–46.
Rabbia M, Guridi Ormazabal M, Staunton H, Veenstra K, Eggenspieler D, Annoussamy M, et al. Stride Velocity 95th Centile Detects Decline in Ambulatory Function Over Shorter Intervals than the 6-Minute Walk Test or North Star Ambulatory Assessment in Duchenne Muscular Dystrophy. J Neuromuscul Dis. 2024;11(3):701–14.
U.S. Food and Drug Administration. Medical Device Development Tool (MDDT) summary of evidence and basis of qualification: Apple Atrial Fibrillation History Feature (U230116) [Internet]. 2024 [cited 2026 May 21]. Available from: https://www.fda.gov/media/178230/download
Apple Inc. Using Apple Watch for arrhythmia detection [Internet]. 2020 [cited 2026 May 22]. Available from: https://www.apple.com/healthcare/docs/site/Apple_Watch_Arrhythmia_Detection.pdf
U.S. Food and Drug Administration. De Novo classification request for Irregular Rhythm Notification Feature: decision summary (DEN180042) [Internet]. 2018 [cited 2026 May 22]. Available from: https://www.accessdata.fda.gov/cdrh_docs/reviews/DEN180042.pdf
Perez MV, Mahaffey KW, Hedlin H, Rumsfeld JS, Garcia A, Ferris T, et al. Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med. 2019;381(20):1909–17.
U.S. Food and Drug Administration. 510(k) premarket notification: Atrial Fibrillation History Feature (K213971) [Internet]. 2022 [cited 2026 May 22]. Available from: https://www.accessdata.fda.gov/cdrh_docs/pdf21/K213971.pdf
Vorpal Technologies K.K. Apple no shinbousaidou rireki program [Apple AFib History Feature]: package insert (Approval No. 30600BZI00010000, 2nd ed.) [Internet]. 2025 [cited 2026 May 22]. Available from: https://www.apple.com/legal/ifu/afhf/afhf-ifu-ja_JP_japan.pdf
Abbott Diabetes Care Inc. FreeStyle Libre Flash Glucose Monitoring System User’s Manual [Internet]. 2017 [cited 2026 Sept 14]. Available from: https://www.accessdata.fda.gov/cdrh_docs/pdf16/P160030C.pdf
Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42(8):1593–603.
U.S. Food and Drug Administration. Summary of safety and effectiveness data (SSED): FreeStyle Libre Flash Glucose Monitoring System (PMA P160030) [Internet]. 2017 [cited 2026 May 22]. Available from: https://www.accessdata.fda.gov/cdrh_docs/pdf16/P160030B.pdf
Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. The Lancet. 2016;388(10057):2254–63.
Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline J-P, Rayman G. Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial. Diabetes Ther. 2017;8(1):55–73.
Alsahli MA, Alalwan A, Aburisheh KH, Alarifi FF, Alshaya HM, Alkholaif AF, et al. Assessing satisfaction, quality of life, and HbA1c changes in type 1 diabetes patients who are using freestyle libre glucose monitoring. J Fam Med Prim Care. 2024;13(6):2367–74.
Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther. 2015;17(11):787–94.
Al Hayek AA, Robert AA, Al Dawish MA. Acceptability of the FreeStyle Libre Flash Glucose Monitoring System: The Experience of Young Patients With Type 1 Diabetes. Clin Med Insights Endocrinol Diabetes. 2020;13:1179551420910122.
Davey RJ, Low C, Jones TW, Fournier PA. Contribution of an Intrinsic Lag of Continuous Glucose Monitoring Systems to Differences in Measured and Actual Glucose Concentrations Changing at Variable Rates in Vitro. J Diabetes Sci Technol. 2010;4(6):1393–9.
Giza S, Kotanidou EP, Tsinopoulou VR, Poukoulidou T, Ntouma S, Galli-Tsinopoulou A. Can Glucose Alarm Fatigue Threaten the Absolute Clinical Benefit of Continuous Glucose Monitoring in Optimal Glucose Management in Children and Adolescents with Type 1 Diabetes? A Narrative Review. Children. 2025;12(12):1668.
Zhang H, Li J, Wang D, Du Q, Zhang J. Impact of continuous glucose monitoring on patient-reported outcomes in adults with type 2 diabetes: a systematic review and meta-analysis. Front Endocrinol. 2026;17:1830980.
Abbott Japan LLC. FreeStyle Libre sensor: package insert (Approval No. 22800BZX00212000, 9th ed.) [Internet]. 2022 [cited 2026 May 22]. Available from: https://www.pmda.go.jp/PmdaSearch/kikiDetail/ResultDataSetPDF/100159_22800BZX00212000_B_01_08
Rochester L, Mazzà C, Mueller A, Caulfield B, McCarthy M, Becker C, et al. A Roadmap to Inform Development, Validation and Approval of Digital Mobility Outcomes: The Mobilise-D Approach. Digit Biomark. 2020;4(Suppl 1):13–27.
Atrsaei A, Corrà MF, Dadashi F, Vila-Chã N, Maia L, Mariani B, et al. Gait speed in clinical and daily living assessments in Parkinson’s disease patients: performance versus capacity. Npj Park Dis. 2021;7(1):24.
Warmerdam E, Hausdorff JM, Atrsaei A, Zhou Y, Mirelman A, Aminian K, et al. Long-term unsupervised mobility assessment in movement disorders. Lancet Neurol. 2020;19(5):462–70.
Mazzà C, Alcock L, Aminian K, Becker C, Bertuletti S, Bonci T, et al. Technical validation of real-world monitoring of gait: a multicentric observational study. BMJ Open. 2021;11(12):e050785.
Micó-Amigo ME, Bonci T, Paraschiv-Ionescu A, Ullrich M, Kirk C, Soltani A, et al. Assessing real-world gait with digital technology? Validation, insights and recommendations from the Mobilise-D consortium. J NeuroEngineering Rehabil. 2023;20(1):78.
Keogh A, Alcock L, Brown P, Buckley E, Brozgol M, Gazit E, et al. Acceptability of wearable devices for measuring mobility remotely: Observations from the Mobilise-D technical validation study. Digit Health. 2023;9:20552076221150745.
Mikolaizak AS, Rochester L, Maetzler W, Sharrack B, Demeyer H, Mazzà C, et al. Connecting real-world digital mobility assessment to clinical outcomes for regulatory and clinical endorsement–the Mobilise-D study protocol. PLOS ONE. 2022;17(10):e0269615.
Viceconti M, Tome M, Dartee W, Knezevic I, Hernandez Penna S, Mazzà C, et al. On the use of wearable sensors as mobility biomarkers in the marketing authorization of new drugs: A regulatory perspective. Front Med. 2022;9:996903.
European Medicines Agency. Letter of support for Mobilise-D digital mobility outcomes as monitoring biomarkers (EMA/234828/2020) [Internet]. 2020 [cited 2026 May 22]. Available from: https://www.ema.europa.eu/en/documents/other/letter-support-mobilise-d-digital-mobility-outcomes-monitoring-biomarkers_en.pdf
European Medicines Agency. Letter of support for Mobilise-D digital mobility outcomes as monitoring biomarkers: follow-up (EMA/CHMP/SAWP/277641/2021) [Internet]. 2021 [cited 2026 May 22]. Available from: https://www.ema.europa.eu/en/documents/other/letter-support-mobilise-d-digital-mobility-outcomes-monitoring-biomarkers-follow_en.pdf
Polhemus A, Delgado-Ortiz L, Brittain G, Chynkiamis N, Salis F, Gaßner H, et al. Walking on common ground: a cross-disciplinary scoping review on the clinical utility of digital mobility outcomes. Npj Digit Med. 2021;4(1):149.
U.S. Food and Drug Administration. 510(k) premarket notification: K110525 — Vitalograph Model 7100 (VitaloJAK) [Internet]. 2011 [cited 2026 May 22]. Available from: https://www.accessdata.fda.gov/cdrh_docs/pdf11/K110525.pdf
McGarvey LP, Birring SS, Morice AH, Dicpinigaitis PV, Pavord ID, Schelfhout J, et al. Efficacy and safety of gefapixant, a P2X3 receptor antagonist, in refractory chronic cough and unexplained chronic cough (COUGH-1 and COUGH-2): results from two double-blind, randomised, parallel-group, placebo-controlled, phase 3 trials. The Lancet. 2022;399(10328):909–23.
Mines D, Bacci E, Nguyen AM, Shaffer S, Smith J, Vernon M. Assessment of inter- and intra-rater reliability of objective cough frequency in patients with chronic cough [abstract]. Eur Respir J. 2019;54(Suppl 63):PA4342.
European Medicines Agency, Committee for Medicinal Products for Human Use. Lyfnua (gefapixant): European public assessment report [Internet]. 2023 [cited 2026 May 21]. Available from: https://www.ema.europa.eu/en/documents/assessment-report/lyfnua-epar-public-assessment-report_en.pdf
Pharmaceuticals and Medical Devices Agency. Review report: Lyfnua tablets 45 mg (gefapixant citrate), MSD K.K. [Internet]. 2021 [cited 2026 May 22]. Available from: https://www.pmda.go.jp/drugs/2022/P20210107001/170050000_30400AMX00008_A100_2.pdf
U.S. Food and Drug Administration, Center for Drug Evaluation and Research. Final summary minutes of the Pulmonary-Allergy Drugs Advisory Committee meeting, November 17, 2023 (gefapixant, NDA 215010) [Internet]. 2023 [cited 2026 May 22]. Available from: https://www.fda.gov/media/176352/download
Schelfhout J, Nguyen AM, Birring SS, Bacci ED, Vernon M, Muccino DR, et al. Validation and Meaningful Change Thresholds for an Objective Cough Frequency Measurement in Chronic Cough. Lung. 2022;200(6):717–24.
International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use. ICH harmonised guideline: guideline for good clinical practice E6(R3) Annex 2, step 4 version [Internet]. 2026 [cited 2026 Aug 23]. Available from: https://database.ich.org/sites/default/files/ICH_E6(R3)_Annex%202_Guideline_Step%204_2026_0603_0.pdf
公開済
投稿日時: 2026-10-06 02:56:24 UTC
公開日時: 2026-10-29 00:24:38 UTC
ライセンス
Copyright(c)2026
Osamu Takenouchi
Kiyoyuki Chinzei
Megumi Furukawa
Tadashi Hosoya
Yuya Imai
Norihiro Kawabata
Takashi Moriya
Yuka Suzuki
Mizuki Yoshida
Takanori Fujita
Toshikazu Fukami
Kenta Hosomi
Keisuke Izumi
Taishiro Kishimoto
Naoki Kubota
Shuta Mitomo
Hideki Oi
Yuichiro Yano
Katsuhiko Yoshimoto
Yoichi Ito M.
Kazumichi Minato
この作品は、Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International Licenseの下でライセンスされています。