プレプリント / バージョン1

Current Landscape of Digital Biomarkers: Device Authorization, Regulatory Qualification, and Clinical-Trial Endpoints

##article.authors##

  • Osamu Takenouchi 株式会社テックドクター
  • Toshikazu Fukami TechDoctor, Inc.
  • Takanori Fujita Tokyo Foundation for Policy Research, Tokyo, Japan
  • Kenta Hosomi Kyowa Kirin Co., Ltd.
  • Keisuke Izumi TechDoctor, Inc.
  • Naoki Kubota Eisai Co., Ltd.
  • Shuta Mitomo CMIC HOLDINGS Co., Ltd.
  • Yuichiro Yano Department of General Medicine, Juntendo University Faculty of Medicine
  • Katsuhiko Yoshimoto Nippon Shinyaku Co., Ltd.
  • Kiyoyuki Chinzei Health Research Institute, National Institute of Advanced Industrial Science and Technology (AIST)
  • Megumi Furukawa Johnson & Johnson Innovative Medicine
  • Tadashi Hosoya Department of Rheumatology, Institute of Science Tokyo, Tokyo, Japan
  • Yuya Imai argenx BV
  • Yoichi Ito M. Data Science Center, Promotion Unit, Institute of Health Science Innovation for Medical Care, Hokkaido University Hospital
  • Norihiro Kawabata Chugai Pharmaceutical Co., Ltd.
  • Takashi Moriya Edwards Lifesciences Japan Ltd.
  • Yuka Suzuki International Affairs, Clinical Research Innovation and Education Center, Tohoku University Hospital
  • Mizuki Yoshida Department of Biometrics & Data Management, Development Japan, Pfizer R&D Japan G.K.
  • Hideki Oi Department of Healthcare Management, Kyorin University School of Medicine
  • Taishiro Kishimoto Hills Joint Research Laboratory for Future Preventive Medicine and Wellness, Keio University School of Medicine
  • Kazumichi Minato TechDoctor, Inc.

DOI:

https://doi.org/10.51094/jxiv.6687

キーワード:

Digital Biomarker、 Medical Device、 Qualification、 Digital Endpoint、 Regulatory Landscape

抄録

Introduction

Digital biomarkers (dBMs), which provide objective measures of individuals’ performance in daily life, are expected to promote patient-centered care, to create new value in routine clinical practice and in treatment, and to be used in clinical trials. The adoption of dBMs can be categorized into three distinct pathways: (i) regulatory authorization as a medical device, (ii) qualification for a specific context of use (CoU), and (iii) adoption as a digital endpoint (dEP) in clinical trials, each of which requires different evidence. The lack of comprehensive overviews of these structural differences has led to increased rework during development. This review aimed to systematically map the structure of this adoption gap.

Methods

This study was conducted as a scoping review in accordance with the PRISMA-ScR guidelines. Drawing on publicly available regulatory information from the US Food and Drug Administration (FDA) and the European Medicines Agency (EMA), public registries, clinical-trial databases, and academic literature, medical devices incorporating dBMs, dBMs qualified or supported by regulatory authorities, and dEPs used in clinical trials were collected and organized.

Results

A total of 225 medical devices had been authorized by the FDA, 10 dBMs had been qualified or supported by the FDA and the EMA, and 65 dEPs had been adopted as primary or secondary endpoints across 20 Phase 3 clinical trials listed in the Digital Medicine Society (DiMe) Library of Digital Endpoints. Maturity markedly differed across disease areas, with many areas having a substantial track record in medical devices but not progressing to qualification or primary endpoints in clinical trials. Comparison of advancing and lagging areas suggests that this maturity gap reflects three conditions: a reliable gold standard in the disease, consensus on the value of continuous measurement, and a clearly defined CoU.

Conclusion

Each of the three adoption pathways carries distinct evidence requirements, and a dBM established in one pathway is not automatically accepted in another. A fit-for-purpose evidence strategy needs to be designed from the earliest stages, anchored in the target CoU and the intended adoption pathway.

利益相反に関する開示

The conflict of interest statement is included in the main text of the manuscript.

ダウンロード *前日までの集計結果を表示します

ダウンロード実績データは、公開の翌日以降に作成されます。

引用文献

Erdemli G, Murphy T, Walinsky S. Regulatory considerations for successful implementation of digital endpoints in clinical trials for drug development. Npj Digit Med. 2025;8(1):142.

Tackney MS, Steele A, Newman J, Fritzsche M-C, Lucivero F, Khadjesari Z, et al. Digital endpoints in clinical trials: emerging themes from a multi-stakeholder Knowledge Exchange event. Trials. 2024;25(1):521.

Di J, Demanuele C, Kettermann A, Karahanoglu FI, Cappelleri JC, Potter A, et al. Considerations to address missing data when deriving clinical trial endpoints from digital health technologies. Contemp Clin Trials. 2022;113:106661.

Babrak LM, Menetski J, Rebhan M, Nisato G, Zinggeler M, Brasier N, et al. Traditional and Digital Biomarkers: Two Worlds Apart? Digit Biomark. 2019;3(2):92–102.

Aboagye NY, Hinchliffe C, Del Din S, Ng W-F, Baker KF, Baker MR. Systematic review: digital biomarkers of fatigue in chronic diseases. Npj Digit Med. 2025;8(1):602.

Bakker JP, Izmailova ES, Clement A, Hoffmann S, Leptak C, Menetski JP, et al. Regulatory Pathways for Qualification and Acceptance of Digital Health Technology-Derived Clinical Trial Endpoints: Considerations for Sponsors. Clin Pharmacol Ther. 2025;117(1):56–72.

Rochester L, Mazzà C, Mueller A, Caulfield B, McCarthy M, Becker C, et al. A Roadmap to Inform Development, Validation and Approval of Digital Mobility Outcomes: The Mobilise-D Approach. Digit Biomark. 2020;4(Suppl 1):13–27.

Manta C, Patrick-Lake B, Goldsack JC. Digital Measures That Matter to Patients: A Framework to Guide the Selection and Development of Digital Measures of Health. Digit Biomark. 2020;4(3):69–77.

Servais L, Strijbos P, Poleur M, Mirea A, Butoianu N, Sansone VA, et al. Evidentiary basis of the first regulatory qualification of a digital primary efficacy endpoint. Sci Rep. 2024;14(1):29681.

Rabbia M, Guridi Ormazabal M, Staunton H, Veenstra K, Eggenspieler D, Annoussamy M, et al. Stride Velocity 95th Centile Detects Decline in Ambulatory Function Over Shorter Intervals than the 6-Minute Walk Test or North Star Ambulatory Assessment in Duchenne Muscular Dystrophy. J Neuromuscul Dis. 2024;11(3):701–14.

Izmailova ES, Wood WA, Liu Q, Zipunnikov V, Bloomfield D, Homsy J, et al. Remote Cardiac Safety Monitoring through the Lens of the FDA Biomarker Qualification Evidentiary Criteria Framework: A Case Study Analysis. Digit Biomark. 2021;5(1):103–13.

Kunkoski E, Saha A, Kam MM. FDA Perspective on the Importance of Digital Health Technologies in Clinical Trials. JAMA Health Forum. 2025;6(4):e250165.

Flora C, Tyler J, Mayer C, Warner DE, Khan SN, Gupta V, et al. High-frequency temperature monitoring for early detection of febrile adverse events in patients with cancer. Cancer Cell. 2021;39(9):1167–8.

Dillenseger A, Weidemann ML, Trentzsch K, Inojosa H, Haase R, Schriefer D, et al. Digital Biomarkers in Multiple Sclerosis. Brain Sci. 2021;11(11):1519.

Sahandi Far M, Eickhoff SB, Goni M, Dukart J. Exploring Test-Retest Reliability and Longitudinal Stability of Digital Biomarkers for Parkinson Disease in the m-Power Data Set: Cohort Study. J Med Internet Res. 2021;23(9):e26608.

Goldsack JC, Dowling AV, Samuelson D, Patrick-Lake B, Clay I. Evaluation, Acceptance, and Qualification of Digital Measures: From Proof of Concept to Endpoint. Digit Biomark. 2021;5(1):53–64.

Godfrey A, Vandendriessche B, Bakker JP, Fitzer‐Attas C, Gujar N, Hobbs M, et al. Fit‐for‐Purpose Biometric Monitoring Technologies: Leveraging the Laboratory Biomarker Experience. Clin Transl Sci. 2021;14(1):62–74.

Garcia J, Mordin MM, Banjara B, Saande C, Keyloun KR, Abel JL, et al. Sensor-based digital health technologies to capture endpoints in recent clinical trials: a scoping review. Npj Digit Med. 2026;9(1):417.

Bakker JP, Barge R, Centra J, Cobb B, Cota C, Guo CC, et al. V3+ extends the V3 framework to ensure user-centricity and scalability of sensor-based digital health technologies. Npj Digit Med. 2025;8(1):51.

Vasudevan S, Saha A, Tarver ME, Patel B. Digital biomarkers: Convergence of digital health technologies and biomarkers. Npj Digit Med. 2022;5(1):36.

Izmailova ES, Demanuele C, McCarthy M. Digital health technology derived measures: Biomarkers or clinical outcome assessments? Clin Transl Sci. 2023;16(7):1113–20.

Coravos A, Doerr M, Goldsack J, Manta C, Shervey M, Woods B, et al. Modernizing and designing evaluation frameworks for connected sensor technologies in medicine. Npj Digit Med. 2020;3(1):37.

Celik Y, Godfrey A. Bringing it all together: Wearable data fusion. Npj Digit Med. 2023;6(1):149.

Daniore P, Nittas V, Haag C, Bernard J, Gonzenbach R, von Wyl V. From wearable sensor data to digital biomarker development: ten lessons learned and a framework proposal. Npj Digit Med. 2024;7(1):161.

Jadoenathmisier KD, Gardarsdottir H, Mol PGM, Pasmooij AMG. Insights from the European Medicines Agency on digital health technology derived endpoints. Drug Discov Today. 2025;30(6):104388.

Perry B, Kehoe L, Swezey T, Le Masne Q, Goldhahn J, Staley A, et al. How Much Evidence Is Enough? Research Sponsor Experiences Seeking Regulatory Acceptance of Digital Health Technology-Derived Endpoints. Digit Biomark. 2023;7(1):45–53.

Perez MV, Mahaffey KW, Hedlin H, Rumsfeld JS, Garcia A, Ferris T, et al. Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med. 2019;381(20):1909–17.

Chang P-C, Wen M-S, Chou C-C, Wang C-C, Hung K-C. Atrial fibrillation detection using ambulatory smartwatch photoplethysmography and validation with simultaneous holter recording. Am Heart J. 2022;247:55–62.

Pepplinkhuizen S, Hoeksema WF, van der Stuijt W, Van Steijn NJ, Winter MM, Wilde AAM, et al. Accuracy and clinical relevance of the single-lead Apple Watch electrocardiogram to identify atrial fibrillation. Cardiovasc Digit Health J. 2022;3(6):S17–22.

Pereira T, Tran N, Gadhoumi K, Pelter MM, Do DH, Lee RJ, et al. Photoplethysmography based atrial fibrillation detection: a review. Npj Digit Med. 2020;3(1):3.

Wang H, Liu B, Zhang H, Zhang Z, Jin Z, Wang H, et al. Photoplethysmography-Based Machine Learning Approaches for Atrial Fibrillation Burden: Algorithm Development and Validation. JMIR Cardio. 2025;9:e78075.

Kim YG, Choi J-I, Oh I-Y, Jeong JH, Lee HS, Choi YY, et al. A Patch-Type Electrocardiography Is Superior to Holter Monitoring for Detecting Paroxysmal Cardiac Arrhythmias. J Korean Med Sci. 2025;40(28):e168.

Kim S, Choi Y, Lee K, Kim S-H, Kim H, Shin S, et al. Comparison of the 11-Day Adhesive ECG Patch Monitor and 24-h Holter Tests to Assess the Response to Antiarrhythmic Drug Therapy in Paroxysmal Atrial Fibrillation. Diagnostics. 2023;13(19):3078.

Foo DHP, Law WC, Wong SC, Tay JS, Ng BHS, Chunggat J, et al. Comparison of adhesive single-lead cardiac ECG patch device with Holter monitoring for detecting atrial fibrillation after an acute cerebral ischemic event: An interim analysis. Eur Heart J. 2023;44(Suppl 1):ehac779.022.

Gao Y, Moayedi Y, Foroutan F, Verma B, Kim B, De Luca E, et al. Remote monitoring of heart failure exacerbations using a smartwatch. Nat Med. 2026;32(3):924–33.

Golbus JR, Gosch K, Birmingham MC, Butler J, Lingvay I, Lanfear DE, et al. Association Between Wearable Device Measured Activity and Patient-Reported Outcomes for Heart Failure. JACC Heart Fail. 2023;11(11):1521–30.

Barrett PM, Komatireddy R, Haaser S, Topol S, Sheard J, Encinas J, et al. Comparison of 24-hour Holter Monitoring with 14-day Novel Adhesive Patch Electrocardiographic Monitoring. Am J Med. 2014;127(1):95.e11-7.

Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42(8):1593–603.

American Diabetes Association Professional Practice Committee. 7. Diabetes Technology: Standards of Care in Diabetes—2026. Diabetes Care. 2026;49(Suppl 1):S150–65.

Bevan A, Ellis G, Eskandarian M, Garrisi D. The Application of Continuous Glucose Monitoring Endpoints in Clinical Research: Analysis of Trends and Review of Challenges. J Diabetes Sci Technol. 2026;20(2):317–24.

U.S. Food and Drug Administration. Diabetes Mellitus: Efficacy Endpoints for Clinical Trials Investigating Antidiabetic Drugs and Biological Products. Guidance for Industry. Draft Guidance [Internet]. 2023 [cited 2026 May 21]. Available from: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/diabetes-mellitus-efficacy-endpoints-clinical-trials-investigating-antidiabetic-drugs-and-biological

European Medicines Agency. Guideline on clinical investigation of medicinal products in the treatment or prevention of diabetes mellitus (CPMP/EWP/1080/00 Rev.2) [Internet]. 2023 [cited 2026 May 21]. Available from: https://www.ema.europa.eu/en/clinical-investigation-medicinal-products-treatment-or-prevention-diabetes-mellitus-scientific-guideline

Berry SE, Valdes AM, Drew DA, Asnicar F, Mazidi M, Wolf J, et al. Human postprandial responses to food and potential for precision nutrition. Nat Med. 2020;26(6):964–73.

Zeevi D, Korem T, Zmora N, Israeli D, Rothschild D, Weinberger A, et al. Personalized Nutrition by Prediction of Glycemic Responses. Cell. 2015;163(5):1079–94.

Childs A, Mayol B, Lasalde-Ramírez JA, Song Y, Sempionatto JR, Gao W. Diving into Sweat: Advances, Challenges, and Future Directions in Wearable Sweat Sensing. ACS Nano. 2024;18(36):24605–16.

Goldstein C, Ghanbari H, Sharma S, Collop N, Namen A, Kirsch DB, et al. Polysomnography validation of SANSA to detect obstructive sleep apnea. Front Neurol. 2025;16:1592690.

Van Braeckel E, Braun AT, Simmonds NJ, Barry PJ, Davies JC, Egan ME, et al. Effects of Elexacaftor/Tezacaftor/Ivacaftor on cough frequency, physical activity patterns, and sleep quality in adolescents and adults with cystic fibrosis. J Cyst Fibros. 2026;25(4):682–8.

Watz H, Troosters T, Beeh KM, Garcia-Aymerich J, Paggiaro P, Molins E, et al. ACTIVATE: the effect of aclidinium/formoterol on hyperinflation, exercise capacity, and physical activity in patients with COPD. Int J Chron Obstruct Pulmon Dis. 2017;12:2545–58.

Nathan SD, Rajicic N, Dudenhofer R, Hussain R, Argula R, Bandyopadhyay D, et al. Inhaled Nitric Oxide in Fibrotic Lung Disease: A Randomized, Double-Blind, Placebo-controlled Trial. Ann Am Thorac Soc. 2024;21(12):1661–9.

Janssen Pharmaceutical K.K. A study of macitentan in Japanese pediatric participants with pulmonary arterial hypertension (NCT05167825) [Internet]. 2021 [cited 2026 May 21]. Available from: https://clinicaltrials.gov/study/NCT05167825

Regeneron Pharmaceuticals. A study to evaluate the effect of dupilumab on exercise capacity in adult patients with asthma (NCT04203797) [Internet]. 2019 [cited 2026 May 21]. Available from: https://clinicaltrials.gov/study/NCT04203797

Sanofi. A phase 4 randomized, double-blind, placebo-controlled study of the effect of dupilumab on sleep disturbance in patients with uncontrolled persistent asthma (MORPHEO) (NCT04502862) [Internet]. 2020 [cited 2026 May 21]. Available from: https://clinicaltrials.gov/study/NCT04502862

U.S. Food and Drug Administration. 510(k) premarket notification: Embrace (K172935) [Internet]. 2018 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K172935

U.S. Food and Drug Administration. 510(k) premarket notification: Embrace (K181861) [Internet]. 2018 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K181861

U.S. Food and Drug Administration. 510(k) premarket notification: EpiMonitor (K232915) [Internet]. 2024 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K232915

U.S. Food and Drug Administration. 510(k) premarket notification: EpiMonitor (K250515) [Internet]. 2025 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K250515

U.S. Food and Drug Administration. 510(k) premarket notification: EpiWatch Monitoring System (K243515) [Internet]. 2025 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K243515

U.S. Food and Drug Administration. 510(k) premarket notification: NightWatch+ US (K243199) [Internet]. 2025 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K243199

U.S. Food and Drug Administration. 510(k) premarket notification: Personal Kinetigraph (PKG) System (K161717) [Internet]. 2016 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K161717

U.S. Food and Drug Administration. 510(k) premarket notification: Personal Kinetigraph (PKG) System Gen 2 Plus (K211887) [Internet]. 2022 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K211887

U.S. Food and Drug Administration. 510(k) premarket notification: Parky App (K220820) [Internet]. 2022 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K220820

U.S. Food and Drug Administration. 510(k) premarket notification: NeuroRPM (K221772) [Internet]. 2023 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K221772

U.S. Food and Drug Administration. 510(k) premarket notification: Rune Labs Tremor Transducer System (K213519) [Internet]. 2022 [cited 2026 May 21]. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K213519

European Medicines Agency. Qualification opinion on stride velocity 95th centile as a secondary endpoint in Duchenne muscular dystrophy measured by a valid and suitable wearable device [Internet]. 2019 [cited 2026 May 21]. Available from: https://www.ema.europa.eu/en/documents/scientific-guideline/qualification-opinion-stride-velocity-95th-centile-secondary-endpoint-duchenne-muscular-dystrophy-measured-valid-and-suitable-wearable-device_en.pdf

European Medicines Agency. Final qualification opinion for stride velocity 95th centile as primary endpoint in studies in ambulatory Duchenne muscular dystrophy studies (EMADOC-1700519818-1127132) [Internet]. 2023 [cited 2026 May 21]. Available from: https://www.ema.europa.eu/en/documents/scientific-guideline/qualification-opinion-stride-velocity-95th-centile-primary-endpoint-studies-ambulatory-duchenne-muscular-dystrophy-studies_en.pdf

Mendell JR, Muntoni F, McDonald CM, Mercuri EM, Ciafaloni E, Komaki H, et al. AAV gene therapy for Duchenne muscular dystrophy: the EMBARK phase 3 randomized trial. Nat Med. 2025;31(1):332–41.

Guglieri M, Campbell C, Deconinck N, De Waele L, Flanigan K, Lorentzos M. Sustained functional improvement with DYNE-251 in males with DMD mutations amenable to exon 51 skipping enrolled in the phase 1/2 DELIVER trial. Neuromuscul Disord. 2025;53(1):105632.

Torous J, Onnela J-P, Keshavan M. New dimensions and new tools to realize the potential of RDoC: digital phenotyping via smartphones and connected devices. Transl Psychiatry. 2017;7(3):e1053.

Insel TR. Digital Phenotyping: Technology for a New Science of Behavior. JAMA. 2017;318(13):1215–6.

Ranjan Y, Rashid Z, Stewart C, Conde P, Begale M, Verbeeck D, et al. RADAR-Base: Open Source Mobile Health Platform for Collecting, Monitoring, and Analyzing Data Using Sensors, Wearables, and Mobile Devices. JMIR MHealth UHealth. 2019;7(8):e11734.

Zulueta J, Piscitello A, Rasic M, Easter R, Babu P, Langenecker SA, et al. Predicting Mood Disturbance Severity with Mobile Phone Keystroke Metadata: A BiAffect Digital Phenotyping Study. J Med Internet Res. 2018;20(7):e241.

Zimmerman M, Ellison W, Young D, Chelminski I, Dalrymple K. How many different ways do patients meet the diagnostic criteria for major depressive disorder? Compr Psychiatry. 2015;56:29–34.

Fried EI, Nesse RM. Depression is not a consistent syndrome: An investigation of unique symptom patterns in the STAR*D study. J Affect Disord. 2015;172:96–102.

Mulinari S. Aligning digital biomarker definitions in psychiatry with the National Institute of Mental Health Research Domain Criteria framework. NPP—Digital Psychiatry Neurosci. 2024;2(1):15.

公開済


投稿日時: 2026-10-06 02:56:41 UTC

公開日時: 2026-11-04 00:07:44 UTC
研究分野
臨床医学