Accuracy Evaluation of an iPad Light Detection and Ranging-based Breathing Motion Measurement System Using a Breathing Motion Phantom
DOI:
https://doi.org/10.51094/jxiv.6569キーワード:
Breathing motion、 Non-contact measurement、 Respiratory function、 Phantom study抄録
Objectives: We aimed to evaluate the accuracy of an iPad light detection and ranging (LiDAR)-based breathing motion measurement system using a breathing motion phantom. The system measured regional breathing motion in the chest and abdominal regions determined using body keypoints, and the measurements were compared with those obtained using an optical motion capture system.
Methods: A breathing motion phantom generated sinusoidal anterior–posterior motion under nine combinations of breathing rates (5, 10, and 15 breaths/min) and peak-to-peak motion ranges (10, 20, and 30 mm). Breathing motion was measured simultaneously using an iPad Pro with a LiDAR-based breathing motion measurement application and an optical motion capture system. Each condition was measured five times in the chest and abdominal regions. The measurement accuracy was assessed using the mean absolute error (MAE), root mean square error (RMSE), and signal-to-noise ratio (SNR).
Results: MAE was 0.39–0.69 mm and 0.33–0.79 mm and RMSE was 0.47–0.88 mm and 0.42–0.95 mm for the chest and abdomen, respectively. The largest errors were observed at 15 breaths/min with a 30-mm peak-to-peak range. SNR was 16.96–25.46 dB for the chest and 16.66–24.26 dB for the abdomen.
Conclusions: The iPad LiDAR-based breathing motion measurement system demonstrated submillimeter measurement errors under the tested breathing conditions. These findings support the feasibility of using an iPad-based LiDAR system for non-contact measurement of regional breathing motions. Further validation in human participants with more physiologically representative breathing patterns and environmental conditions is required.
利益相反に関する開示
The authors are affiliated with a joint research department established by JA Kyosai Research Institute and the Institute of Science Tokyo. AN is an unpaid director of Vitalizar Inc. The BREVI iPad application was provided free of charge by Vitalizar Inc.ダウンロード *前日までの集計結果を表示します
引用文献
Mendes LPDS, Vieira DSR, et al.: Influence of posture, sex, and age on breathing pattern and chest wall motion in healthy subjects, Braz J Phys Ther. 2020; 24(3): 240–248.
Bradley H, Esformes J.: Breathing pattern disorders and functional movement, Int J Sports Phys Ther. 2014; 9(1): 28–39.
Jung S-H, Hwang U-J, et al.: Does mobilisation of the thoracic spine using mechanical massage affect diaphragmatic excursion in individuals with thoracic hyperkyphosis? J Back Musculoskelet Rehabil. 2022; 35(3): 517–523.
Ristovski A, Kapeleti M, et al.: Acute effects of diaphragmatic breathing on trunk and shoulder mobility and pulmonary function in healthy young adults, J Funct Morphol Kinesiol. 2025; 10(3): 325.
Tassani S, Chaves P, et al.: Breathing, postural stability, and psychological health: a study to explore triangular links, Front Bioeng Biotechnol. 2024; 12.
Courtney R, Cohen M, et al.: Comparison of the manual assessment of respiratory motion (MARM) and the Hi Lo breathing assessment in determining a simulated breathing pattern, Int J Osteopath Med. 2009; 12(3): 86–91.
Kiesel K, Rhodes T, et al.: Development of a screening protocol to identify individuals with dysfunctional breathing, Int J Sports Phys Ther. 2017; 12(5): 774–786.
Bockenhauer SE, Chen H, et al.: Measuring thoracic excursion: Reliability of the cloth tape measure technique, J Am Osteopath Assoc. 2007; 107(5): 191–196.
Mohan V, Dzulkifli NH, et al.: Intrarater reliability of chest expansion using cloth tape measure technique, Bangladesh J Med Sci. 2012; 11(4): 307–311.
Mohan V, Paungmali A, et al.: Respiratory characteristics of individuals with non-specific low back pain: A cross-sectional study, Nurs Health Sci. 2018; 20(2): 224–230.
Cala SJ, Kenyon CM, et al.: Chest wall and lung volume estimation by optical reflectance motion analysis, J Appl Physiol. 1996; 81(6): 2680–2689.
Koseki T, Kakizaki F, et al.: Effect of forward head posture on thoracic shape and respiratory function, J Phys Ther Sci. 2019; 31(1): 63–68.
Aramaki Y, Kakizaki F, et al.: Effects of the posterior pelvic tilt sitting posture on thoracic morphology and respiratory function, J Phys Ther Sci. 2021; 33(2): 118–124.
Nozoe M, Mase K, Tsutou A.: Regional chest wall volume changes during various breathing maneuvers in normal men, J Jpn Phys Ther Assoc. 2011; 14(1): 12–18.
Massaroni C, Nicolo A, et al.: Contactless methods for measuring respiratory rate: A review, IEEE Sens J. 2021; 21(11): 12821–12839.
Aoki H, Nakamura H.: Non-contact respiration measurement during exercise tolerance test by using kinect sensor, Sports (Basel). 2018; 6(1): 23.
Kempfle J, Van Laerhoven K.: Breathing in-depth: A parametrization study on RGB-D respiration extraction methods, Front Comput Sci. 2021; 3.
Nocera A, Senigagliesi L, et al.: Development and validation of a frequency-based quality index for contactless respiratory volume analysis, IEEE Sens J. 2026; 26(3): 5062–5076.
Nocera A, Chebbaki A, et al.: Extraction of the respiratory volume signal with an Intel RealSense D455 depth camera: A qualitative study, in 2025 IEEE International Conference on Consumer Technology-Europe (ICCT-Europe). IEEE. 2025: i–iv.
Benetazzo F, Freddi A, et al.: Respiratory rate detection algorithm based on RGB-D camera: theoretical background and experimental results, Healthc Technol Lett. 2014; 1(3): 81–86.
Xie W, Xu C, et al.: DeepBreath: Breathing exercise assessment with a depth camera, Proc ACM Interact Mob Wearable Ubiquitous Technol. 2024; 8(3): 1–26.
Sun C, Li W, et al.: An unobtrusive and non-contact method for respiratory measurement with respiratory region detecting algorithm based on depth images, IEEE Access. 2019; 7: 8300–8315.
Yasukawa M, Noyori SS, et al.: Non-contact 3D measurement method for evaluating thoracoabdominal motion in respiratory assessment, in 2022 61st Annual Conference of the Society of Instrument and Control Engineers (SICE). IEEE. 2022: 815–820.
Spears SDJ, Lester T, et al.: Comparative evaluation of Artec Leo hand-held scanner and iPad Pro for 3D scanning of cervical and craniofacial data: assessing precision, accuracy, and user experience, 3D Print Med. 2024; 10(1).
Kyriazidis I, Berner JE, et al.: 3D breast scanning in plastic surgery utilizing free iPhone LiDAR application: Evaluation, potential, and limitations, Aesthet Surg J. 2025; 45(4): NP99–NP104.
Capaldi DP, Hirata E, et al.: A smartphone-based motion monitoring system for surface guided radiation therapy, Adv Radiat Oncol. 2026; 11(3): 101970.
Vitalizar Inc.: BREVI (In Japanese). https://vitalizar.net/brevi/spec (Accessed on September 21, 2026)
Apple Inc.: Detecting human body poses in images — Apple developer documentation. https://developer.apple.com/documentation/vision/detecting-human-body-poses-in-images (Accessed on September 21, 2026)
ダウンロード
公開済
投稿日時: 2026-09-22 06:22:52 UTC
公開日時: 2026-10-23 07:26:39 UTC
ライセンス
Copyright(c)2026
野寄, 修平
二村, 昭元
この作品は、Creative Commons Attribution-NoDerivatives 4.0 International Licenseの下でライセンスされています。